Abstract

A robust tandem HPLC method coupling size-exclusion (Shodex Asahipak GS-320HQ) and reversed phase (Vydac 218TP54) columns with ultraviolet detection was developed for quantitative determination of interferon-gamma (IFN-gamma) in Chinese hamster ovary cell culture supernatant. The 2D-HPLC system was linked up by a 6-port 2-position low hold-up volume switch valve. Compared to a commercial ELISA kit for IFN-gamma, the coupled column LC approach was able to detect and quantify soluble IFN-gamma, regardless of the glycoprotein's molecular/conformational variability and sample background. Each LC-LC analysis took 90 minutes inclusive of column regeneration. The relative standard deviation of measurements (n = 5) was less than 3%. The limit of detection (LOD) was determined to be 0.35 microg IFN-gamma.