Abstract

A simple preparation scheme is described for the quantitative analysis of a magnesium niobate sample using slurry introduction axially viewed inductively coupled plasma optical emission spectrometry. Relationships between the stability of slurries and the conditions, such as particle size, pH, dispersant and amount of dispersant, were investigated experimentally. The MgNb(2)O(6) slurry sample was prepared by adding the dispersant sodium polyacrylate and agitation in an ultrasonic bath to ensure good dispersion. Under optimization of pH and amount of dispersant, an analysis of minor and trace impurities (Ba, Ca, Cr, Cu, Fe, Mn, Ni, Pb) in magnesium niobate was accomplished. Applying a paired t test, we showed that the results were in agreement at a 95% confidence level with the reference values obtained by a fusion method for a magnesium niobate sample, which verified that the calibration curves could be established by aqueous standards. Analytical results demonstrate that the factors that affected the accuracy of determination for MgNb(2)O(6) are mainly the particle size of the sample and the stability of slurry.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.