Abstract

A gradient normal-phase high-performance liquid chromatography (NP-HPLC) method has been developed for the determination of impurities in α-cyclodextrin. The method is selective for six linear sugars and several late-eluting unidentified impurities, as well as for α-, β- and γ-cyclodextrins. The HPLC system utilizes mobile phases consisting of acetonitrile–water (A, 64:36; B, 35:65), a 5-μm YMC Polyamine II column (250×4.6 mm I.D.), and pulsed amperometric detection. Experiments were conducted to determine the optimal mobile phase concentrations, establish the precision of the method, and demonstrate linearity and determine relative response factors for select sugars. The method was applied to select lots of commercial-grade α-cyclodextrin from two different manufacturers. The method is novel in that it offers gradient elution of strongly-retained impurities at ambient column temperatures and pulsed amperometric detection.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.