The development and characterisation of a new biosensor for hydroperoxides is described, which is obtained by combining an oxygen gas diffusion amperometric electrode and two immobilized enzymes (peroxidase and tyrosinase) working in parallel and competing for the same substrate (catechol). The response of the biosensor to several hydroperoxides was investigated (LOD=0.5·10 −4 M for hydrogen peroxide). It was experimentally found that the biosensor is able to respond also to aqueous solutions of ionic peroxides (LOD=0.2·10 −4 M for potassium peroxidisulphate). The biosensor was applied to the determination of the hydrogen peroxide content of pharmaceutical products, i.e. aqueous disinfectant solutions (RSD% ⩽0.5; recoveries by standard addition method between 96.0 and 98.5%).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.