Abstract

ABSTRACTAn instrument for prompt gamma-ray activation analysis (PGAA) at the NIST Center for Neutron Research has proven useful for the measurement of hydrogen and other elements in a variety of materials. The sample is irradiated by a beam of low energy neutrons. Gamma-rays emitted by atomic nuclei upon neutron capture are measured and elemental concentrations determined by comparison with appropriate standards. The detection limit for hydrogen is < 5 mg/kg in most materials, and 2 mg/kg for hydrogen measured in silicon. The instrument has been used to measure hydrogen mass fractions of < 100 mg/kg in high purity germanium, and < 10 mg/kg in quartz. More recently PGAA has been used to measure hydrogen in 1 μm thick porous thin films on a silicon substrate, and in crystals of silicon carbide and cerium aluminate.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.