Abstract

A simple, rapid, precise and eco-friendly ion chromatography (IC) method for the determination of hippuric acid (HA) in human urine was proposed in this paper. The separation was carried out an anion exchange column with 2.0 mmol L −1 Na 2CO 3 + 2.0 mmol L −1 NaHCO 3 as mobile phase at the flow-rate 0.7 mL min −1. A suppressed conductivity detector was used and the detection limit was 1.0 μg L −1(S/N = 3) for hippuric acid. The analysis time for one run was 30 min under the optimized IC condition. The recovery of hippuric acid was 93.2–98.0% while the relative standard deviation (RSD) was 1.4–2.3% by seven measurements.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.