Abstract

Measuring hemoglobin mass (Hbmass) using the carbon monoxide (CO) bolus rebreathing method is frequently used in research but has yet to be widely used in the clinical practice. The estimation of an adequate CO bolus may be difficult in patients with unknown Hbmass. In the present pilot study, a progressive inhalation technique for CO that leads to a linear individual adjusted COHb increase was evaluated. Sixteen healthy test subjects participated in the study (preliminary investigation: six; main study: ten). The reliability and validity of the new method were evaluated using multiple measurements of Hbmass with and without a defined blood donation and compared to a CO bolus method. The participants inhaled a CO‐air mixture (CO concentration: 1500 ppm) for a specific breathing duration. The CO uptake and COHb change were determined simultaneously. The typical error (reliability) in the repeated measurements was 2.4% (CI ± 4.7). The mean difference between the new method and the bolus method was 34 g (±41; P = 0.026). The measured hemoglobin loss in 490 mL of blood was 74 g (±35), and the calculated hemoglobin loss was 77 g (±4) (mean difference 3 g ± 34; P = 0.820). The new method was reliable and valid in a proof of concept study with healthy subjects. The total amount of CO and as a result the COHb increase is individually adjustable. Future studies in clinical settings are needed to determine if the method could be used in disease‐specific pathologies associated with changes in Hbmass.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call