Haloacetic acids (HAAs) were determined in different water samples by a new, fast and simple analysis method based on enrichment of 50-ml water samples at pH 1.8 by solid-phase extraction (SPE) followed by liquid chromatography (LC) separation and electrospray ionization mass spectrometric detection in the negative ionization mode. Deprotonated (M-H) − haloacetates and decarboxylated (M-COOH) − ions were detected. Different polymeric SPE sorbents were tested, and LiChrolut EN was found to be the best material for the extraction. Complete LC separation of all compounds could only be achieved by ion-pair chromatography using triethylamine as volatile ion-pairing reagent. The detection limits were in the low μg/l range. High μg/l concentration levels for the chlorinated and brominated haloacetates were found in drinking water from a drinking water treatment plant in Barcelona, and the corresponding tap water. In swimming pool water samples from Catalonia mg/l levels and in surface river water from Portugal μg/l values were detected. These results confirm other recent reports on the ubiquitous occurrence of HAAs in aqueous environments.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.