Abstract
An accurate quantitative determination of halide ions X (X = Cl, Br, I) in aqueous solution by total reflection X-ray fluorescence (TXRF) is not possible using the traditional acidic internal standards. In general, the standard solutions are highly acidic (e.g., Ga(NO3)3 in HNO3) to avoid precipitation of hydroxides of the standard element and to obtain a stable and reliable standard. In acidic solutions, dissolved halide salts can exchange their cation for a proton. The resulting volatile HX compounds can evaporate during the drying procedure of the TXRF sample preparation. In this technical note, we show that an alkaline Cu(NH3)4(NO3)2 standard can be used for the determination of chlorine, bromine and iodine without facing problems of HX evaporation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.