Abstract
This paper suggests a sensitive reversed-phase gradient HPLC method combined with fluorescence detection that has been developed, optimized and tested via the quantitative analysis of authentic biological material in an effort to determine and subsequently compare the total content of glycosaminoglycans (GAGs) in various collagen-based biomaterials intended for medical application.The proposed analytical method enabled the identification and separation of the GAGs present from the other components in the samples using commonly-available laboratory equipment; moreover, the very low detection limit of the method permits the determination of GAGs even for very small samples.This study describes the development of the method, including the isolation and processing of the collagen samples prior to HPLC analysis and the optimal parameters applied during the chromatographic analysis. The application of the method in laboratory practice is documented by means of several examples of the determination of GAGs employing both commercial standards and real collagen samples isolated from various animal tissues.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.