Abstract

A simple and sensitive procedure for the determination of gentamicin is presented, based on the use of the peroxyoxalate chemiluminescent (PO-CL) system in presence of imidazole as a catalyst. The gentamicin has to be previously derivatized with o-phthaladehyde (OPA) in order to obtain a fluorophore, which participates in the PO reaction, producing a CL emission proportional to the gentamicin concentration. The method is developed by using a particular flow-injection analysis (FIA) manifold, employing sodium dodecyl sulfate (SDS) micellar medium as a carrier in order to avoid the degradation of PO in water. The optimization of the instrumental and chemical variables affecting the CL reaction was rigorously carried out by using experimental design methodology. The method has been successfully applied to pharmaceutical formulations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.