Abstract
The determination of gallium traces is carried out on the hanging drop mercury electrode in solutions of low ionic strength and in absence of complexing agents. At pH=3.2, where stable solutions are obtained, concentrations of 0.2 μg Ga l−1 are determinated with a standard deviation of 0.96% and a detection limit of 4 ng Ga l−1.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.