Abstract

A micellar electrokinetic chromatography method (MEKC) is developed and validated for the analysis of fluticasone propionate (FP) in nasal sprays. The MEKC method is performed on a fused-silica capillary (50 mum i.d.; effective length, 40 cm). The background electrolyte consists of 25 mM borate and 25 mM anionic detergent SDS solution at pH 9. The capillary temperature is maintained at 35 degrees C, and the applied voltage is 20 kV. The injection is performed using the hydrodynamic mode at 50 mbar for 6 s with detection at 238 nm. The method is linear in the range of 2-80 mug/mL (r(2) = 0.9956). The specificity and stability-indicating capability are proven through forced degradation studies inclusive by mass spectrometry, which also shows that there is no interference of the excipients. The limit of detection and limit of quantitation are 0.56 and 2 mug/mL, respectively. Moreover, method validation demonstrates acceptable results for accuracy, precision, and robustness. The proposed method was successfully applied for the quantitative analysis of FP nasal sprays, and the results were compared to a validated reversed-phase liquid chromatographic method, showing non-significant difference (P > 0.05).

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call