Abstract

Both fluoxetine (FLX) and its N-demethylated metabolite, norfluoxetine (NFLX), have been reported to be potent serotonin-reuptake inhibitors. A sensitive and reliable method that allows simultaneous quantification of their plasma levels would be valuable and was developed in this work. The procedure included extraction of FLX and NFLX from plasma, fluorescence derivatization with 4-(N-chloroformylmethyl-N-methyl) amino-7-nitro-2,1,3-benzoxadiazole (NBD-COCl), separation of the derivatives on an octadecylsilica column with acetonitrile-water (55:45,v/v) as mobile phase and fluorescence detection with emission at 537 nm and excitation at 478 nm. The calibration curves were linear for FLX and NFLX concentration over the range of 10-1000 nM (r = 0.9992 and r = 0.9997) and the limits of quantitation were 10 nM in 100 micro L of plasma. Precision of intra- and inter-day RSD of less than 12% and accuracy of intra- and inter-day RE within -6.0-13% were achieved. The method described was applied to analysis of the plasma samples from rats treated with FLX hydrochloride and to the pharmacokinetic study.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.