Abstract

Fluconazole concentrations in biological fluids were determined by high-performance gas chromatography. A simple extraction procedure with chloroform, under basic conditions and after the addition of UK-47,265 as the internal standard and with no evaporation stage, was carried out prior to analysis. A solid injector and a 15-m capillary column, coated with a nonpolar phase and connected to a nitrogen-selective detector that afforded an excellent selectivity and sensitivity, constituted the gas chromatographic system. The duration of each analysis was <4min and the minimum detectable serum concentration was 50 ng/mL. In five patients undergoing chronic peritoneal dialysis, the mean serum concentrations ± SD at 1, 6, and 48h after the intraperitoneal administration of a single dose of fluconazole were, respectively, 325 ± 75, 928 ± 159, and 607 ± 80ng/mL.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.