Abstract
Determination of flavonoid markers quercetin, hesperetin, and chrysin, found in north Iranian citrus honey samples, was carried out by solid phase extraction (SPE) and isocratic liquid chromatographic separation using central composite design. Optimum conditions for SPE were achieved using 10 mL methanol/water (13:87, v/v, pH = 7) as the washing solvent and 4 mL methanol for elution. Good clean-up and high recovery >90% were observed for all analytes. The use of water/ACN/THF/AcOH (54:36:5:5, v/v) was found to serve as the optimum mobile phase composition and allowed for the separation of analytes from endogenous compounds present in honey. SPE parameters, such as maximum loading capacity and breakthrough volume, were also determined for each analyte. Limit of detection, linear range, recovery, repeatability of retention times, and peak heights were 3.11 × 10−8–4.44 × 10−8 g g−1, 0.50–50.0 μg mL−1 (R 2 > 0.99), 90.7–96.9%, 3.0–3.6%, and 1.0–2.6%, respectively. Precision of the overall analytical procedure, estimated by five replicate measurements for quercetin, hesperetin and chrysin in citrus honey, as well as the relative standard deviations were 4.3%, 3.8%, and 5.5%, respectively.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.