Abstract

A method based on ultra high performance liquid chromatography-tandem mass spectrometry (UHPLC-MS/MS) was established for the determination of fipronil and its metabolites in feeds. Sodium chloride was added to the samples after the ultrasonic extraction of pure water and acetonitrile, and then cleaned up by the QuEChERS method. The target analytes were separated on a Waters HSS T3 C18 column using methanol and water as mobile phases with gradient elution, and detected by a MS/MS system under negative electro-spry ionization (ESI-) and the multiple reaction monitoring (MRM) mode. The limits of detection (LODs) were 0.05 μg/kg, and the limits of quantification (LOQs) were 0.2 μg/kg. The linear correlation coefficients were more than 0.9998. The average recoveries ranged from 92.3% to 105.4% with RSDs no more than 2.3%. This method is suitable for the identification of fipronil and its metabolites in feeds because of its simplicity and high sensitivity.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.