Abstract

We have developed a "sandwich"-type time-resolved immunofluorometric assay (IFMA) for fetal hemoglobin (HbF) in hemolysates from adults and newborns, amniotic fluid, and plasma, based on a polyclonal and a monoclonal antibody against human fetal hemoglobin. Microtiter wells are coated with polyclonal capture antibody, and the gamma-chain-specific monoclonal tracer antibody is labeled with a europium chelate. In a simple and fast assay procedure, prediluted hemolysates are incubated in the microtiter wells first with capture antibody for 1 h and, after washing, for 1 h with tracer antibody. The wells are washed and the fluorescence of europium is measured. The mean analytical recovery is 102% and results by IFMA agreed well with values obtained by high-performance liquid chromatography. The analytical range of IFMA is large and well suited for clinical purposes. The detection limit of the assay is 0.2 microgram/L and the measuring range extends to 500 micrograms/L.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.