Abstract

Felodipine in plasma was extracted with toluene and determined by automated capillary gas chromatography with electron capture detection. Undesirable oxidation of the dihydropyridine derivative in the injector and in the column was avoided by the use of glass materials with the inner surfaces treated by high-temperature silylation. The day-to-day reproducibility of the method was represented by a relative standard deviation (RSD) of 5% for a felodipine concentration of 25 nmol/l. The minimum determinable concentration (giving better than 15% RSD) was 1-2 nmol/l (0.4-0.8 ng/ml).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.