Abstract

ABSTRACTA sorption balance is an instrument used to measure vapor uptake in a sample at controlled temperature and relative humidity. It is most commonly used to determine equilibrium values (sorption isotherms), but is also used for kinetic measurements of transport coefficients. Such measurements can be affected by the external mass transfer resistance in the gas phase around the sample. This paper presents a method to determine the external mass transfer coefficient for a given flow geometry using a water saturated sample, including corrections for temperature changes from evaporative cooling, which is found to have considerable effect on the calculated constant.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.