Abstract

Abstract A method is presented for the determination of ethylene chlorohydrin and ethylene oxide residues in plastic and rubber surgical equipment. Samples are extracted with p-xylene and the desired constituents are isolated by column chromatography. Ethylene oxide is converted to ethylene chlorohydrin on a Celite column. Sweep co-distillation is then employed for further purification, if necessary, before gas-liquid chromatographic analysis. The lower detection limit is approximately 25 ng. Recoveries of 3.5’50 ppm chlorohydrin from 20’35 g samples range from 64 to 100%, depending on the absorbing capacity of the material being analyzed and the condition of the adsorbent.