Abstract

To develop a method for determining epichlorohydrin in the workplace air by gas chromatograph-electron capture detector (GC-ECD). Epichlorohydrin in the workplace air was collected by activated charcoal tubes, desorbed using acetone, and analyzed by GC-ECD. A good linearity was obtained in the range of 1.0-50 µg/mL (r=0.999 7). The detection limit was 0.012 µg/ml, while the recovery rate was 88.1% and relative standard deviation ranged from 1.11% to 3.57%. The samples could be stored for seven days at room temperature. This method effectively eliminates the interferences of alkanes on determination of epichlorohydrin and improves the sensitivity by 1 to 2 orders of magnitude, which can solve the problem of detection limit above standard in GBZ/T 160.58-2004.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.