Abstract

A method for the determination of endogenous agmatine in rat plasma was developed by isotope dilution-gas chromatography-negative chemical ionization mass spectrometry (GC-NCI/MS). The plasma samples were analyzed after protein precipitation, evaporation, derivatization by hexafluoroacetone (HFAA), and clean-up on a Florisil SPE column. The GC-MS analysis utilized stable isotope d8-agmatine as internal standard. The samples after treatme were tested by negative chemical ionization with selected ion monitoring (SIM) which was set at m/z 492 (molecular ion of agmatine) and m/z 500 (molecular ion of internal standard). The limit of detection (LOD) of agmatine standard solution was 0.005 7 ng/mL. The calibration curve of the agmatine spiked in rat plasma showed a good linear relationship at the range of 1.14-57.0 ng/mL (r = 0.997). The recoveries of agmatine spiked in rat plasma ranged from 92.3% to 109.8%. Inter-day and intra-day precisions were less than 15%. The average concentration level of agmatine in rat plasma was (22 +/- 9) ng/mL, and there was no significant difference between male and female SD rats (p > 0.05). The method is high sensitive and specific, and can be used for the determination of endogenous agmatine in plasma. It provides a strong support for the subsequent research of agmatine.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call