Abstract

It is shown that direct analysis of viscous organic liquids (medical and industrial adhesives, polymeric flocculant) by arc atomic emission spectrometry using aqueous standard solutions for constructing calibration dependences is impossible due to the matrix effect. This limitation is eliminated when the samples are
 diluted by at least in an order of magnitude, which leads to an increase in the limits of detection. A simplified scheme of acid mineralization of samples directly on the electrode (50 μL of nitric acid per 10 μL of sample) is proposed. The absence of the influence of the matrix is confirmed by the calculation of plasma
 parameters and comparison with the results obtained after traditional acid decomposition. The proposed method allows in a short time to obtain information on the content of a set of elements (Ag, Al, B, Ba, Be, Bi, Ca, Cd, Co, Cr, Cu, Fe, Mg, Mn, Ni, P, Pb, Ti, V, Zn) from hundredths of mg/kg in viscous organic liquids.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call