Abstract
AbstractAccurate determination of the flexion-extension axis of the elbow affects the outcome of implant replacement. The current study proposes an automated approach capable of determining the FE axis based on a stack of axial computer tomographic (CT) imaging slices of the distal humerus. The core of the algorithm consists of an original technique employing control polygon deformation used to approximate the segmented outer cortical bone points with closed B-Splines, followed by curvaturebased and least squares fitting methods for determination of the two relevant geometric centers. The new approach was validated against a conventional voxelbased FE axis determination procedure involving marching cubes algorithm.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.