Abstract

Epoxyeicosatrienoic acids (EETs) are cytochrome P-450 metabolites of arachidonic acid involved in the regulation of vascular tone. The method of microbore column high-performance liquid chromatography with fluorescence detection was developed to determine 14,15-EET, 11, 12-EET, and the mixture of 8,9-EET and 5,6-EET. Tridecanoic acid (TA) was used as an internal standard. EETs were reacted with 2-(2, 3-naphthalimino)ethyl trifluoromethanesulfonate (NT) to form highly fluorescent derivatives. A C(18) microbore column and a water-acetonitrile mobile phase were used for separation. Samples were excited at 259 nm, and the fluorescence was detected at 395 nm. The overall recoveries were 88% for EETs and 40% for TA. EETs were detected in concentrations as low as 2 pg (signal-to-noise ratio = 3). The method was used to determine the EET production from endothelial cells (ECs). Bradykinin and methacholine (10(-6) M) stimulated an increase in the production of EETs by ECs two- and fivefold, respectively. This sensitive method may be used for determination of EETs at low concentrations normally detected in complex biological samples.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.