Abstract

Diacetyl is an important flavoring compound in many foods, especially in beer. In the present study, we developed and validated a new precolumn derivatization HPLC-UV method for the determination of diacetyl using 4-(2,3-dimethyl-6-quinoxalinyl)-1,2-benzenediamine as a novel derivatizing reagent. After derivatization with the reagent at a pH value 4.0 at ambient temperature for 10 min, diacetyl was analyzed on an ODS column and detected at 254 nm. The results show that the correlation coefficient of the method is 0.9991 in the range of 0.10 to 100.0 μM diacetyl, and the limit of detection is 0.02 μM. The method was further evaluated in the analysis of beer samples with the recoveries ranging from 94.4 to 102.6% and RSDs from 1.36 to 3.33%. The concentrations of diacetyl in 8 beer samples were determined in the range of 0.19 to 0.42 μM. The method established in this study may be well suitable for the determination of diacetyl in beer.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.