Abstract

ABSTRACTThe determination of damaged starch and diastatic activity in flour was studied using a flow‐injection analysis (FIA) biosensor system. The system consisted of an oxygen electrode and an immobilized enzyme column containing purified glucoamylase and glucose oxidase immobilized on activated aminopropyl glass beads. The biosensor system has an optimum pH between 6.5 and 7.5 and an optimum temperature of 35°C for glucose measurement. The response of the FIA biosensor was linear up to 1.000 g/L of glucose with a lower detection limit of 0.025 g/L. Each assay took about 20 min, and the system showed good reproducibility (r = 0.998, n = 8). When applied to the measurement of damaged starch and diastatic activity in wheat flour, the results obtained agreed with those obtained using the conventional methods of measurement. This biosensor system is a rapid practical alternative for the measurement of damaged starch and diastatic activity in wheat flour.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call