Abstract

A new method for the analysis of cycloserine (4-amino-3-isoxazolidinone, CYC) in rat microdialysis samples has been developed. This method consists of derivatizing the CYC with benzoyl chloride, which transforms primary amines into highly stable derivatives. An attractive feature of this method was that the derivatization reaction is straightforward and can be completed within 10 min. The formed derivative, in contrast to the non-derivatized analyte, exhibited increased chromatographic retention and decreased matrix effects resulting from the co-elution of other components using reversed-phase liquid chromatography and on-line switching. Detection on a quadrupole-linear ion trap mass spectrometer (AB3200 Q-Trap) was performed using electrospray tandem mass spectrometry in multiple reaction monitoring mode. Various derivatization parameters were optimized in order to improve chromatographic separation and minimize ion suppression. In particular, the benzoylation reaction was improved to enhance the reproducibility and sensitivity of the chromatographic method. The transition m/z 207.1 → 105.1 was acquired to monitor the CYC derivatization products. The method was fully validated for its sensitivity, selectivity, matrix effect and stability. A good linearity over the selected range (r > 0.99, range = 22-2200 mg/L), as well as accuracy and precision within ±7% of the target values, was obtained. The assay described herein was successfully applied to quantitatively measure CYC in the lung and blood of anesthetized rats.