Abstract

The application of reversed-phase gradient high-performance liquid chromatography with UV detection to the determination of cortisol and cortisone in 24-h urine samples is described. The method employs Sep-pak C18 cartridges for the part-purification and concentration of the corticosteroids, with sample enrichment at the head of an HPLC pre-column and separation using water/acetonitrile gradient. The internal standard is 6alpha-methylprednisolone. Measurement of both cortisone and cortisol provides further information on adrenocortical function. 24-hour excretion rate data from normal subjects are reported.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.