Abstract

Electrothermal atomic absorption spectrometry (ETA-AAS) and atomic emission spectrometry (ETA-AES) have been applied to the determination of copper in powdered milk. A homogeneous dispersion procedure for the preparation of the milk powder is described which was found to be simple, rapid and less susceptible to contamination than dry ashing or wet digestion methods. Both ETA-AAS and ETA-AES techniques were found to provide satisfactory results using conventional tube wall atomisation only when the method of standard additions was employed. The application of graphite probe atomisation in ETA-AES and ETA-AAS allowed the development of direct methods for the determination of copper in milk powder using aqueous standard calibration curves. The accuracy of the probe ETA-AAS method was confirmed using new reference materials prepared by the EEC Community Bureau of Reference. Acceptable agreement was obtained for the other procedures using a commercial milk powder sample with a copper content of 6.0μg g−1.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.