Abstract
ABSTRACTAn online flow injection procedure for the preconcentration of copper at µg L−1 levels was developed using a novel chelating resin and flame atomic absorption spectrometry for the analysis of industrial water. Amberlite XAD-2 functionalized with dithizone resin was packed in a minicolumn for flow injection. The pH, sample and eluent flow rates, and eluent concentration were optimized to ensure maximum recovery of Cu(II). The effect of concomittant ions on Cu (II) sorption was also investigated. The limit of detection and enrichment factor for a 180 s preconcentration time were 76 and 0.6 µg L−1, respectively. The resin offered precision of 1.0% for seven successive determinations of 100 µg L−1 Cu(II). The resin was used for more than 300 sorption and desorption cycles without appreciable loss of efficiency. The accuracy of the preconcentration procedure was confirmed by fortified recovery studies in industrial water with NIST certified copper nitrate with recovery exceeding 96%. Validation was performed by the analysis of a standard reference material.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.