Abstract

In this study, an effective Fenton digestion procedure based on UV irradiation was used to facilitate copper determination in human blood serum by flame atomic absorption spectrometry. The method was based on the degradation of blood serum samples using free hydroxyl radicals that were generated from the mixture of iron magnetic nanoparticles and hydrogen peroxide under UV irradiation. The digestion procedure was carried out using a simple, multi-functional lab-made UV irradiation reactor. Under the optimal experimental conditions, detection and quantification limits (LOD and LOQ) of the digestion system for copper were found to be 0.15 and 0.51 mg/L, respectively. Different blood serum samples were taken through the optimum procedure to validate the digestion method. Satisfactory percent recoveries (93–107%) were obtained with high repeatability for replicate digestions. This promising digestion method is simple, uses very low sample and reagent volumes, and can therefore be considered as an alternative procedure to microwave digestion.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.