Abstract

The analytical conditions for the electrothermal atomic absorption spectrometric determination of copper, manganese, nickel, and cobalt in different types of baby foods using slurried samples are presented. Suspensions were prepared in a medium containing 0.1% w/v Triton X-100, 30% v/v concentrated hydrogen peroxide, 1% v/v concentrated nitric acid, 1% w/v ammonium dihydrogenphosphate (only for copper atomization), and one drop of silicon antifoam. The slurries were homogenized by using a potter and were introduced directly into the furnace. The amount of analyte extracted to the liquid phase was investigated. The graphite furnace conditions were optimized for each element. No matrix effect was observed, and calibration could be carried out using aqueous standard solutions. The detection limits were 3.4, 2.1, 5.4, and 3.7 pg for copper, manganese, nickel, and cobalt, respectively. The reliability of the procedures was checked by statistically comparing the results obtained with those found by using a previous microwave-oven mineralization stage and by analyzing several certified reference materials.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.