Abstract
A sensitive, simple and selective high-performance liquid chromatographic (HPLC) method was developed for the determination of ciprofloxacin in biological fluids. The method is based on the reaction between the drug and 4-chloro-7-nitrobenzofurazan (NBD-Cl) in borate buffer of pH 9.0 to yield a highly fluorescent derivative that is measured at 535 nm after excitation at 464 nm. The calibration curves were linear over the concentration ranges of 25–3000 and 50–3000 ng·mL−1 for plasma and urine, respectively. The mean recovery of ciprofloxacin from plasma and urine was 98.37% and 98.40%, respectively. The method was found to be sensitive, precise, accurate, and reproducible. All of the validation parameters were within the acceptance range.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.