Abstract

A simple analytical procedure, making use of the differential volatilities of the CrIII and CrVI complexes with trifluoroacetylacetone, was developed for the determination of CrVI and total Cr in river water. It was based on a modification of the electrothermal atomic absorption spectrometric method with the addition of ammonium trifluoroacetylacetonate before starting the thermal programme. The analytical procedures developed was found to be applicable to the determination of 2–300 µg l–1 of CrVI in the presence of up to 300 µg l–1 of CrIII. The precision was found to vary from 3.8 to 1.4% for 2.5–25 µg l–1 of CrVI, respectively. The applicability of the method developed to the analysis of river water samples was established by using parallel-method comparison, and no statistically significant difference between the results obtained and those of an ion-exchange resin method was observed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.