Abstract

A reliable and rapid method is described for the determination of chromium in alcoholic beverages by electrothermal atomic absorption spectrometry (ETAAS) after optimization of the temperature-time program for the graphite furnace. Chromium in wines was determined directly without sample pretreatment. Other samples were first mineralized with HNO3 and V2O5. The method was validated, and its analytical characteristics were checked. The detection limit was 1 pg (for a sample volume of 10 microL). Accuracy was 99.20 +/- 3.80% for wines and 98.90 +/- 2.00% for other beverages. Precision ranged from 3.50 to 4.50%. The proposed method was applied to 124 samples of alcoholic beverages frequently consumed in Spain. Mean values obtained (microgram/L) were 20.28 for wines, 8.75 for beer, 7.85 for cider, 14.60 for brandy, 7.50 for rum, 10.25 for whisky, 8.80 for gin, 9.25 for vodka, 24.45 for anisette, and 10.08 for liquors. The advantages of the proposed method (rapidity, sensitivity) and versatility) make it useful for routine analyses and for estimations of dietary intake of chromium.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.