Abstract
A fast, simple, and sensitive sample preparation procedure based on dispersive liquid–liquid microextraction (DLLME) is proposed for the determination of cholesterol in food samples using isocratic reverse phase high performance liquid chromatography (RP-HPLC) and UV detection. The influence of several important parameters on extraction efficiency of cholesterol was evaluated. Under optimized conditions, a linear relationship was obtained between the peak area and the concentration of cholesterol in the range of 0.03–10 μg l −1. The detection and quantification limits were 0.01 and 0.03 μg l −1, respectively. Intra-day and inter-day precisions for the analysis of cholesterol were in the range of 1.0–3.1%. The applicability of the proposed method was demonstrated by analyzing cholesterol in milk, egg yolk and olive oil.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.