Abstract

An HPLC method for simultaneous determination of chlorogenic acid and mangiferin in original medicinal materials and decoction pieces of Pyrrosiae Folium was developed. The assay was performed on a Diamonsil C18 (4.6 mm x 250 mm, 5 microm) column eluted with a mobile phase consisted of acetonitrile and 0.5% phosphoric acid solution in gradient elution at a flow rate of 1.0 mL x min(-1). The column temperature was set at 25 degrees C. The detection wavelength was 320 nm. The results showed that The linear ranges of chlorogenic acid and mangiferin were 5.2-130 mg x L(-1) (r = 0.9999) and 1.2-18 microg x mL(-1) (r = 0.9999), and the average recoveries (n=6) were 97.9% (RSD 1.9%) and 99.6% (RSD 2.9%), respectively. The method was simple, reproducible and valid. It can be used for quality evaluation and control of original medicinal materials and decoction pieces of Pyrrosiae Folium.