Abstract
Abstract A single‐channel flow injection analysis (FIA) system for the pseudo‐titration of chloride in soils was developed using a tubular silver ion selective electrode with a homogeneous crystalline membrane of Ag2S as potentio‐metric detector. This system allows the determination of chloride in soil extracts from 1 to 100 ppm. The results obtained for 15 soil extracts are in good agreement with those provided by the classical procedures with sampling rates varying from 60 to 160 determinations per hour with good precision (relative standard deviations lower than 4.8%).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.