Abstract

A direct potentiometric method involving titration against a standard volumetric silver nitrate solution using a silver electrode to detect the end point was evaluated for the determination of chloride in infant formula and adult/pediatric nutritional formula. It was assessed for compliance against AOAC Standard Method Performance Requirements (SMPR®) established by the AOAC Stakeholder Panel on Infant Formula and Adult Nutritionals (SPIFAN). A single-laboratory validation (SLV) study was conducted as a first step in the process to validate the method. In this SLV, 17 SPIFAN matrixes representing a range of infant formula and adult nutritional products were evaluated for their chloride content. The analytical range was found to be between 1.4 and 1060 mg/100 g reconstituted product or ready-to-feed (RTF) liquid. The LOQ was estimated as 1.4 mg/100 g. Method repeatability was between 0.03 and 1.60% in the range of 20 to 167 mg/100 g RTF, and intermediate precision was between 0.09 and 2.77% in the same range. Recovery values based on spiking experiments at two different levels of chloride ranged from 99.0 to 103% for 15 different SPIFAN products. Evaluation of trueness was performed on National Institute of Standards and Technology Standard Reference Material 1849a (Infant/Adult Nutritional Formula) and showed 97.2% of the theoretical value, with no bias at the 95% confidence level. Based on the results of the SLV, the method met the SMPR and was approved as a First Action method by the AOAC Expert Review Panel on Infant Formula and Adult Nutritionals on March 17, 2015.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call