Abstract

This paper describes the evaluation of a method for determination of Cd and Pb in xanthan gum samples by Graphite Furnace Atomic Absorption Spectrometry (GF AAS) using NH4H2PO4 as the chemical modifier. The sample preparation was performed using a reflux system adapted in the digestion tubes. With this system it was possible to increase the temperature of the digester block above the boiling point of the reaction medium, preventing loss of analyte and excessive evaporation of acids during heating. Samples were digested with HNO3 for 3 h in a digester block at 220 oC. The limits of detection for Cd and Pb were 2.2 and 33.8 ng g-1, respectively. The RSDs for both analytes were, on average, lower than 5.0% and accuracy was verified by recovery tests, yielding values in the 83-100% range.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.