Abstract

The determination of carisoprodol and its metabolite meprobamate in oral fluid using solid-phase extraction and liquid chromatography with tandem mass spectral detection (LC-MS-MS) and its application to authentic specimens is described. The method employs collection of oral fluid with the Quantisal device, extraction using cation exchange/hydrophobic solid-phase columns, and LC-MS-MS in positive ion electrospray mode. The method was fully validated using various parameters, including selectivity, linearity, accuracy, intra-day and inter-day imprecision, drug recovery from the collection pad, limit of quantitation and matrix effects. The method was applied to both routine research specimens and an authentic specimen taken from an individual prescribed a daily dose of 350 mg carisoprodol following surgery.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.