Abstract

A flow injection spectrophotometric method for the determination of carbofuran in commercial pesticide formulations was developed. The determination involves on-line hydrolysis of the extracted carbofuran at room temperature with sodium hydroxide. The resulting carbofuran-phenol is coupled with diazotized 4-aminobenzoic acid in order to achieve an appropriate selectivity and sensitivity for the spectrophotometric measurements. The calibration curve is linear over the range 1–10 mg litre−1 of carbofuran. The proposed method has a detection limit of 0.15 mg litre−1 of carbofuran and a sample frequency of 120 injections per hour. The relative standard deviation of six independent determinations of a sample containing 1 mg litre−1 carbofuran was 0.6%. The suitability of the proposed procedure for the determination of carbofuran in commercial pesticide formulations was studied. The procedure provides results comparable to those obtained by liquid chromatographic analysis. © 2000 Society of Chemical Industry

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.