Abstract

The use of a tungsten coil as a platform has been evaluated for the determination of carbide-forming metals, Cr and Ba. The methodology developed has been used for Cr and Ba determinations in HPS CRM, Trace Metals in Drinking Water, and NIST SRM 1643b, Trace Elements in Water. Good agreement was found between the results obtained and the certified values for Cr in both reference materials and for Ba in the NIST SRM. Barium results were approximately 14% lower than the certified results for the HPS CRM. This inaccuracy was ascribed to the presence of considerable amounts of Ca in this particular material. Precision of the measurements, 3σ relative standard deviation as percentage, were 2.5 and 5.8 for Cr and Ba, respectively. Characteristic masses, in picograms, were 1.70 for Cr and 19.0 for Ba. These results compare favorably with the literature values for Cr but not for Ba.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.