Abstract

The remote performance of thyroid function blood tests is complicated because it requires blood collection. To compare TSH and free thyroxine (FT4) levels between capillary and venous blood and assess the adequacy of measuring each value in capillary blood. This prospective intervention study was conducted at Ito Hospital and was based on the clinical research method. The participants were 5 healthy female volunteers and 50 patients (41 females and 9 males) between the ages of 23 and 81 years. To measure TSH and FT4 levels in capillary and venous blood, a digital immunoassay (d-IA) method capable of measuring trace samples was used. Chemiluminescence measurements were used as controls. Values obtained for each assay system were compared using Spearman's correlation analysis. Capillary blood was collected using an autologous device (TAP II; not approved in Japan). Capillary plasma volume obtained using TAP II was 125 µL or more in 26 cases, 25 µL to 124 µL in 24 cases, and less than 25 µL in 5 cases. Strong correlations were noted in the TSH and FT4 levels between capillary and venous blood, with correlation coefficients of rs = 0.99 and rs = 0.97, respectively. Capillary TSH and FT4 levels strongly correlate with venous blood values. Trace samples can be used in high-precision d-IA methods. These results may promote telemedicine in assessing thyroid function.