Abstract
An electrochemical vapor generation (EcHG) system to determine the cadmium content using atomic fluorescence spectrometry has been established. The EcHG system uses two serially-connected graphite tubes as two electrolysis cells and an AC to DC power wall plug adapter with the constant voltage mode as a power supply. The system did not have an ion exchange membrane and only needed a pump to transport the feed solution and waste. Under these optimized conditions, the detection limit (3σ) and the relative standard deviation were 0.05 ng mL-1 and 3.2%, respectively. The EcHG efficiency was 38.4 ± 2.2%. The proposed technique has been successfully applied to determine the cadmium in three certified reference materials for environmental water and two drinking water samples.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Analytical sciences : the international journal of the Japan Society for Analytical Chemistry
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.