Abstract
A reduction current is obtained when an aqueous solution of cadmium and 5-fluorouracil (5-FU) is subjected to linear sweep cathodic stripping voltammetry. This current is attributed to the reduction of cadmium-5-FU complex ions which adsorbed onto the hanging mercury drop electrode. The resulting peak current is characterized with respect to supporting electrolytes, pH, accumulation potential, scan rate, possible interferences and other variables. A linear relationship between the peak current and cadmium concentration was obtained over the range 5 × 10−9–1 × 10−7 mol dm−3 with correlation coefficient of 0.9973. A detection limit of 9.0 × 10−11 mol dm−3 was obtained with standard deviation of 1.9 %. The method was successffuly applied for the determination of cadmium in seawater.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.