Abstract

The determination of trace metal contents directly from a solid sample is a trend in modern atomic spectrometry. The aim of this study was to develop an analytical method for the routine determination of Cd in lichens using solid sampling graphite furnace atomic absorption spectrometry (SS-GF-AAS). For the determination of Cd, the temperature program of the graphite furnace was optimized using a mixed matrix modifier (Pd + Mg (NO3)2 + Triton X-100). The limit of detection and the limit of quantification were 0.9μg/kg and 3μg/kg, respectively. The analytical method for Cd determination in the plant matrix was verified by the analysis of certified reference materials of lichens, seaweed, and rye grass. The developed procedure was applied to the study of Cd distribution in thalli of Usnea antarctica lichen from James Ross Island, Antarctica. The SS-GF-AAS analytical method is particularly suited for use in environmental studies and plant physiology (the microanalysis of anatomical structures).

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call