Abstract
A volatile Cd species (presumed to be the hydride) was generated from aqueous solutions by merging sample and tetrahydroborate reductant in a continuous-flow system. The gaseous analyte was transferred onto the inner wall of a graphite tube furnace for in situ preconcentration at 200 degrees C. Calibration was achieved via the method of standard additions. An absolute detection limit (3 sigma blank) of 10 pg was obtained using KBH4 as reductant. The precision of the determination was 12% (RSD) for a Cd concentration of 0.2 ng ml-1 using KBH4. The method was successfully applied to the determination of Cd in several certified environmental reference materials (soil, sediment, sea-water, biological samples) following pressure digestion of the samples by microwave heating in a mixture of acids.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.